Function: Split
Protection level: IP65 for display instrument, IP68 for probe
Probe installation: according to range and probe selection
product details
Selling Units: Single item
Single package size: : 39*26.5*16.5cm
Single gross weight:1.500 KG
English Display Interface
2、After power-on display of the meter, long press the set button (SET) for 2s to enter in the main menu.
3、Input #1 probe height value to “reference zero point 1” which is arranged in the structural chart of the menu as shown in attached table (probe height is the distance from probe emitting surface to tank bottom or pool bottom).
4、Input #2 probe height value to “reference zero point 2” which is arranged in the structural chart of the menu as shown in attached table 1.
5、Set “4mA corresponding difference value” and “20mA corresponding difference value”
4mA corresponding difference value: output 4mA when the difference value is equal to the set value.
20mA corresponding difference value: output 20mA when the difference value is equal to the set value.
The “4mA corresponding difference value” and “20mA corresponding difference value” are situated in the structural chart of the menu as shown in attached table.
I. Purpose
Ultrasonic level difference meter (material level, liquid level) is used to measure the liquid level or material level difference between two containers.
It is most commonly used to measure the water level difference in front and rear of coarse and fine racks at the water inlet of sewage treatment plant, calculate the water level difference and start up the back dragging cleaner to take out rubbish.
Minimum of measuring range: it is the value of distance between the reference plane to the position, which is positive when the minimum of measuring range is above the reference plane and negative when the minimum of measuring range is below the reference plane. The output current is 4mA when the liquid level is at such position.
Maximum of measuring range: it is the value of distance between the reference plane to the position, which is positive when the maximum of measuring range is above the reference plane and negative when the maximum of measuring range is below the reference plane. The output current is 20mA when the liquid level is at such position.
(6)The installed equipment must be grounded truthfully and independently and shall not share the public grounding with electrical cabinet or meter box.
(7)The signal line shall not be wired in the same trunking with the power line and it shall be installed independently through metal tube or far away from the power line. If the signal line is not installed through tube independently, it shall be kept at least 1m away from the power line.
Function | Split type |
Normal measuring range | 5m, 10m, 15m, 20m |
Special measuring range | 30m, 40m, 50m, 60m |
Measuring accuracy | 0.5%-1.0% |
Resolution ratio | 3mm or 0.1% (whichever is greater) |
Display | |
Analog output | 2nd-circuit 4~20mA/ 510Ω load |
Relay output | 4th-circuit AC 250V/ 8A or DC 30V/ 5A state programmable (optional) |
Power supply | 220V AC+15% 50Hz or 24VDC 120m Optional |
Ambient temperature | Display instrument (transmitter): -20~+60ºC Probe (sensor): -20~+80℃ |
Communication | 485 communication (optional) |
IP grade | Display instrument: IP65, probe: IP68 |
Probe cable | Standard 10m, it can reach 200 maximum (it is required to shield all electromagnetic interference) |
Probe installation | Subject to the measuring range and type of probe |
Standard separate-type ultrasonic level meter:
Object Picture Structural Drawing
Thread M48×2 or G2 Sensor Thread M60×2 or G2 Sensor
4.2 Installation Guide
4.2.1 Understand terminology
①Measuring range: the meaning of measuring range is very important for meter type selection. Please refer to the diagrams below.
Product application:
Mainly used in sewage plant, water plant, mine, water conservancy, power plant, chemical plant, steel plant
The monitoring of waterlogging and water, sewage quality, urban pipeline and rainfall station in the construction of smart city project; applicable to the monitoring of rainfall point of existing water conservancy projects, river water and rain monitoring, reservoir water and rain monitoring, flow station, irrigation area, open channel monitoring, etc.; automatic monitoring of water plant, on-line water quality monitoring and secondary water supply monitoring of smart water project; suitable for intelligent water project It is used for the monitoring of collapse and landslide points, debris flow and surface water level of geological disaster projects
TOP