Ultrasonic open channel flowmeter for channels without weir and groove

Model: Ultrasonic open channel flowmeter without weir channel

Protection level: IP65 for display instrument, IP68 for probe

Probe material: The standard configuration is ABS, and anti-corrosion materials should be used in corrosive environments.

product details

一、Packaging & Delivery

Selling Units:  Single item

Single package size: : 39*26.5*16.5cm

Single gross weight:3.00 KG

二、The interface of ultrasonic level meter under operating mode is as follows:

 

 

       English Display Interface

 

1、 Key function:  

Our products can be debugged through three keys on the panel. Then, the measured value is showed by LED.

               

      ◇Enter to menu                       ◇Move the cursor

◇Confirm menu                          ◇Select the menu

◇Confirm parameter calibration          ◇parameter calibration

2、After switched on it, press the SET key for two second and enter the first level menu.

3、The height value of probe input to the “reference zero”.

As the reference zero, please check the figure menu structure.

( Height of probe is the distance from the launch of probe to the edge of weir and flume).

4、“4mA Flow value” and “20mA Flow value”

4mA flow value:output is 4mA when the instantaneous flow equal to this value.

20mA flow value: output is 20mA when the instantaneous flow equal to this value.

The location of "4mA flow value" and "20mA flow value" in the            menu please check Schedule II menu chart.

5、Select the type of weir and flume, you must consider these: the size of flow and dynamics of water in the channel, the formation of free flow or not.
Maximum flow is less than 40 l/s(144 t/h), suggestion is right-angled triangular Weir.

Max flow is more than 40 l/s , the suggestion is parshall flume.

Upstream channel is short and the max flow is more than 40 l/s, the suggestion is rectangular Weir.

Before measurement, you should calibrate the reference zero point which is the distance from the probe to zero point of weir slot level.

(The factory default is parshall flume.)

 

①     Triangular Weir

 As triangular weir, please follow the following instructions:

“9 weir and flume type”---“1 triangular weir”-----“1 work status” then choose the “start”.

For “2 triangular weir angle”, please choose the actual angle of instrument, then the instrument can be calculated the corresponding flow automatically according to water level.

 

② Rectangular Weir

As rectangular weir, please follow the following instructions:

“9 weir and flume type”---“1 rectangular weir”-----“1 work status” then choose the “start”, and choose “0.25m,0.5m,0.75m,1m, Non-standard channels” from menu of “2 standard channel”, then the instrument can be calculated the corresponding flow automatically according to water level.

 

③. Trapezoidal Weir

As trapezoidal weir, please follow the following instructions:

“9 weir and flume type”---“1 trapezoidal weir”-----“1 work status” then choose the “start”, and input actual weir cage width in the menu of “2 weir cage width”, then the instrument can be calculated the corresponding flow automatically according to water level.

④ Parshall flume

As parshall flume, please follow the following instructions:

“9 weir flume type”---“1 parshall flume”-----“1 work status” then choose the “start”. The flow formula of parshall flume: Q=Chan, according to the throat width (b), find the mechanic coefficient(c) and index(n) from the table 2 of the parshall flow formula and input to the menu of “9 weir type”---“1 parshall flume”-----“2 mechanic coefficient c”-----“3 index n”. then the instrument can be calculated the corresponding flow automatically according to water level.

三、Technical data

Function

Split   type


Range

0.1l/s~99999.99 m³/h

Cumulative Flow

biggest:4290000000.00 m³

Max range for level

3m

Accuracy for level

0.5%

Resolution

3mm or 0.1%(bigger)

Display

Chinese LED

Accuracy   for Flow

Standard weir:1~5%

(meet the requirements of national standard weir and   channels)

Non-standard weir: 10~30%。

Analog   output

(Four-wire   system) 4~20mA/750Ωload

Delay output

(optional)two   groups :AC 220V/ 8A or   DC 24V/ 5A,

Power supply

Standard:24VDC 100mA;optional: 220V AC+15%  50Hz。

Power supply

(optional)12VDC、battery、solar

Environment

Temperature

LED:-20~+60℃,Probe: -20~+80℃

Environmental pressures

Standard atmospheric pressure

Environment humidity

≤90%RH, Non-condensing

Process temperature

-20~80℃;

Process pressures

Standard atmospheric pressure

Communication

Optional:485,232,MODBUS protocol

Protection Grade

LED:IP67    Probe: IP68


Cable Probe


Standard:10m,

The longest:100m

Probe installation dimension

M78x2mm thread + matching nuts

Probe

Material

Standard: ABS

Teflon material in corrosive   environment


Power Consumption

Split type:

Power supply:24V,

No relay: 100mA

Channel 1 of Relay: 120mA;

Channel 2 of Relay: 145mA;

Channel 3 of Relay: 170mA;

Channel 4 of Relay: 190mA;

The specific power is as follows:

No relay:  24×100mA=2.4W;

Channel 1 of Relay: 24×120mA=2.9W;

Channel 2 of Relay: 24×145mA=3.5W;

Channel 3 of Relay: 24×170mA=4.1W;

Channel 4 of Relay: 24×190mA=4.6W;

Channel 3 of Relay: 24×170mA=4.1W;

Channel 4 of Relay: 24×190mA=4.6W;

Integrated type (Four-wire   system)

Power supply:24V,

No relay: 80mA

Channel 1 of Relay: 105mA;

Channel 2 of Relay: 130mA;

The specific power is as follows:

No relay:  24×80mA=1.9W;

Channel 1 of Relay: 24×105mA=2.5W;

Channel 2 of Relay: 24×130mA=3.1W;

 四、Installation

◆Integrated type

  ◆Split type

 163780859968734.jpg               163780859967122.jpg

 Figure4.1.1 Outline                  Figure 4.1.2 Structure of flow meter

 

The display parts of split type should be installed the indoors with good ventilation and non-corrosive gases. Instrument is wall mounting. If the condition of indoor is bad or it must be hung out the door, installation of instrument must be in the protective box to avoid the sun and rain.


V. Field cases:

7.jpg

Product application:

 

Mainly used in sewage plant, water plant, mine, water conservancy, power plant, chemical plant, steel plant

The monitoring of waterlogging and water, sewage quality, urban pipeline and rainfall station in the construction of smart city project; applicable to the monitoring of rainfall point of existing water conservancy projects, river water and rain monitoring, reservoir water and rain monitoring, flow station, irrigation area, open channel monitoring, etc.; automatic monitoring of water plant, on-line water quality monitoring and secondary water supply monitoring of smart water project; suitable for intelligent water project It is used for the monitoring of collapse and landslide points, debris flow and surface water level of geological disaster projects


VI. Manufacturer:

TOP